Accuracy, validity, and limits
The short answer: FreeMoCap's accuracy is not a single number. It depends on which pose estimation backend is running, what movement is being recorded, and how the cameras and lighting were set up. The rest of this page is the longer, more useful answer, based on a formal validation study that compared FreeMoCap against marker-based motion capture across gait, balance, and prosthetic alignment.
"The reported metrics of accuracy are not fixed properties of the software package or system. The reported metrics reflect what that system was able to achieve under the specific conditions that it was in."
That line, from the validation study itself, is worth internalizing before reading the numbers below. A number like "30 mm joint trajectory error" describes one system, one camera setup, one pose estimation backend, one task. Change any of those and the number changes with it.
What was validated
A six-camera FreeMoCap setup, built from six webcams costing about $20 each, was compared against a marker-based reference system (Qualisys) across three domains: treadmill gait at increasing walking speeds, static balance (the Modified Clinical Test of Sensory Interaction on Balance, CTSIB-M), and gait changes from controlled alignment adjustments in a transfemoral prosthesis user. Three pose estimation backends were compared throughout: MediaPipe, RTMPose, and ViTPose, plus a custom-trained DeepLabCut model for the prosthetics case.
What the results show
Gait. Joint trajectory errors were generally under 30 mm and sagittal joint angle errors under 5°, in line with ranges reported across other markerless validation studies. The hip, usually one of the hardest joints to track accurately, came in under 25 mm and under 4.5° of flexion/extension error, better than the 30-50 mm and ~11° commonly reported elsewhere. Spatiotemporal gait parameters (stride length, stride duration, and similar) showed strong agreement with the reference; agreement was lowest for swing-phase duration (ICC ≈ 0.90). Errors were largest during early stance, the loading-response phase where the knee flexes quickly to absorb impact, movement that's inherently harder for any video-based system to track, and worth noting that even the marker-based reference isn't perfectly clean here: soft tissue movement over bone introduces its own error during high-impact phases.
Balance. This was, as far as the study's authors could determine, the first markerless validation of the CTSIB-M. Using the MediaPipe backend, center-of-mass path length tracked the marker-based reference closely (ICC = 0.985) and preserved the expected progressive increase in sway across harder balance conditions (removing visual or proprioceptive feedback). RTMPose and ViTPose performed considerably worse on this task, not because of a fundamental accuracy problem, but because both showed higher frame-to-frame landmark jitter at rest, which MediaPipe's built-in smoothing filter suppresses and the other two backends don't.
Prosthetic alignment. Standard, off-the-shelf pose estimation models, trained mostly on able-bodied people, failed to reliably track a transfemoral prosthesis user, whose limb doesn't look like what those models were trained on. Training a custom DeepLabCut model on the prosthetic limb substantially improved tracking, and the resulting system correctly tracked the direction and rough magnitude of leg-length, knee and ankle angle, and foot-progression-angle changes across alignment adjustments, changes clinically relevant even at 5-10 mm. It did not reliably track pelvic obliquity or minimum toe clearance. This case is the clearest demonstration that markerless motion capture is not one fixed accuracy level: swapping in a population-specific model changed the outcome from "doesn't work for this person" to "works well enough to detect real alignment changes."
Backend choice changes the outcome
The three pose estimation backends have different failure modes, not just different accuracy levels:
- MediaPipe (runs without a GPU; FreeMoCap's current V2 pipeline defaults to RTMPose instead) applies temporal smoothing that introduces slight lag during fast movements, visible as speed-dependent deviations in gait kinematics, but that same smoothing is what made it the only backend sensitive enough for the balance study.
- RTMPose was the most consistent backend across the gait study, with the lowest ankle-kinematics error, and showed none of MediaPipe's speed-dependent artifacts.
- ViTPose showed a small, systematic scaling bias (about 2-4% larger than the reference, concentrated in the vertical axis), plausibly a consequence of its vision-transformer architecture lacking the spatial assumptions built into MediaPipe's and RTMPose's CNN-based models, though that explanation hasn't been confirmed. It is also not one of SkellyTracker's built-in detectors (MediaPipe, RTMPose, YOLOX, ArUco, and ChArUco), so running it today means integrating it yourself.
None of these is strictly "more accurate" than the others. Which one to use depends on the task; see choosing a tracking model.
Known limitations
- Temporal resolution. Validation data was collected at 30 Hz, which limits timing measurements to roughly 33 ms increments. Fine-grained timing differences may not be visible at this sampling rate.
- Accuracy is downstream of pose estimation. FreeMoCap's 3D reconstruction is only as good as the 2D keypoints it's given. Occlusion, unusual body shapes, or a pose estimation model that struggles with a particular population degrades results no matter how good the camera setup is.
- Validated conditions are specific. These results come from treadmill gait, static (not dynamic) balance, and one prosthesis user. They don't guarantee the same accuracy for, say, overground sprinting, a sport-specific movement, or a different assistive device. Broader validation across more populations and movement types is ongoing work for the field, not something this system has already covered.
- Capture conditions matter. Camera count, placement, lighting, and occlusion all directly affect reconstruction quality. See optimize your capture space before assuming a result you got doesn't match what's described here.
Where this leaves you
FreeMoCap is not a replacement for marker-based motion capture, which remains the gold standard for precision. What the validation work shows is that a low-cost, open source, markerless system can get close enough to be useful for real biomechanical questions, provided you understand which backend you're using and what it's actually been validated on. Treat this page as a starting point for what to expect, and the glossary and data model pages as where to go to understand exactly what numbers you're getting back.
Source: Cherian, A. Open-source development and validation of a low-cost markerless system for quantitative motion analysis. PhD